Key result
Mitral valve repair using the Medtronic Profile 3D annuloplasty ring yielded 3-year survival rates of 97.1% for isolated procedures and 92.4% for combined procedures (p=0.137).
Why the study?
Does mitral valve repair using the Medtronic Profile 3D annuloplasty ring provide good clinical outcomes in patients with severe degenerative mitral regurgitation?
Observational (n=200)
No
Does mitral valve repair using the Medtronic Profile 3D annuloplasty ring provide good clinical outcomes in patients with severe degenerative mitral regurgitation?
The Medtronic Profile 3D annuloplasty ring provides excellent early and midterm survival and freedom from reoperation in patients with severe degenerative mitral regurgitation.
Supports Profile 3D ring use in degenerative MR; leaves open randomized comparisons to other annuloplasty devices.
Background Various devices have been proposed for ring annuloplasty in patients with degenerative mitral valve disease. This study reports for the first time midterm results with the rigid three-dimensional Medtronic Profile 3D (Medtronic, Minneapolis, Minnesota, United States) annuloplasty ring. Methods Between June 2009 and June 2011, 200 patients (mean age 61 ± 13 years, 70% male) with severe degenerative mitral regurgitation underwent mitral valve repair using the Medtronic Profile 3D annuloplasty ring. A total of 106 patients (53.0%) underwent isolated mitral valve repair and 94 patients (47.0%) underwent a concomitant procedure such as coronary artery bypass grafting (n = 21), tricuspid valve surgery (n = 49), AF ablation (n = 17), and aortic valve surgery (n = 13). The follow-up is 94.5% complete (mean 2.5 ± 0.5 years). Results Thirty-day mortality was 1.5%. Survival at 3 years was 97.1 ± 1.6% for isolated procedures and 92.4 ± 2.8% for combined procedures (p = 0.137). Freedom from mitral valve-related reoperation at 3 years was 97.1 ± 1.7% for isolated procedures and 95.5 ± 2.2% for combined procedures (p = 0.561). Seven patients (3.5%) required a mitral valve-related reoperation. Two of these reoperations were required for endocarditis, two for ring dehiscence, one for progression of the native disease (flail leaflet), one for leaflet suture dehiscence, and one for persistent systolic anterior motion. Conclusion The three-dimensional Medtronic Profile 3D annuloplasty ring is suitable for mitral valve repair for degenerative diseases. This saddle-shaped annuloplasty device provides excellent early results with a very good functional outcome at midterm either in isolated or combined procedures.
No takes yet. Share an insight, caveat, or question.
Ratschiller et al. (2015) conducted an observational in severe degenerative mitral regurgitation (n=200). Medtronic Profile 3D annuloplasty ring (isolated procedures) vs. Combined procedures was evaluated on Survival at 3 years (p=0.137). Mitral valve repair using the Medtronic Profile 3D annuloplasty ring yielded 3-year survival rates of 97.1% for isolated procedures and 92.4% for combined procedures (p=0.137).