Key result
Increasing the stator bore radius of the continuous-flow total artificial heart decreased rotor power, shear stress, and blood residence time while increasing net blood flow rate.
Why the study?
The study was conducted to determine the impact of the stator-bearing bore radius on the journal-bearing hydraulic performance in a continuous-flow total artificial heart while satisfying geometric design constraints.
Does increasing stator bore radius improve journal-bearing hydraulic performance in the computational model of the continuous-flow total artificial heart?
Does increasing stator bore radius improve journal-bearing hydraulic performance in the computational model of the continuous-flow total artificial heart?
Computational fluid dynamics modeling demonstrates that increasing the stator bore radius of the continuous-flow total artificial heart improves journal-bearing hydraulic performance.
Improves modeled hydraulic performance in continuous-flow TAH; leaves open translation to clinical devices.
Cleveland Clinic’s continuous-flow total artificial heart (CFTAH) is a double-ended centrifugal blood pump that has a single rotating assembly with an embedded magnet, which is axially and radially suspended by a balance of magnetic and hydrodynamic forces. The key to the radial suspension is a radial offset between the stator bearing bore and the magnet’s steel laminations. This offset applies a radial magnetic force, which is balanced by a hydrodynamic force as the rotating assembly moves to a “force-balanced” radial position. The journal-bearing blood passage is a narrow flow path between the left and right impellers. The intent of this study was to determine the impact of the stator-bearing bore radius on the journal-bearing hydraulic performance while satisfying the geometric design constraints imposed by the pump and motor configuration. Electromagnetic forces on the journal bearing were calculated using the ANSYS EMAG program, Version 18 (ANSYS, Canonsburg, PA). ANSYS CFX Version 19.2 was then used to model the journal-bearing flow paths of the most recent design of the CFTAH. A transient, moving mesh approach was used to locate the steady state, force-balanced position of the rotating assembly. The blood was modeled as a non-Newtonian fluid. The computational fluid dynamics simulations showed that by increasing stator bore radius, rotor power, stator wall average shear stress, and blood residence time in journal-bearing decrease, while blood net flow rate through the bearing increases. The results were used to select a new bearing design that provides an improved performance compared with the baseline design. The performance of the new CFTAH-bearing design will be confirmed through upcoming in vitro and in vivo testing.
No takes yet. Share an insight, caveat, or question.
Khelghatibana et al. (2021) studied this question. Increasing stator bore radius vs. Baseline design was evaluated on Journal-bearing hydraulic performance (rotor power, stator wall average shear stress, blood residence time, and blood net flow rate). Increasing the stator bore radius of the continuous-flow total artificial heart decreased rotor power, shear stress, and blood residence time while increasing net blood flow rate.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: