Key result
After improvement on the left chamber design of the auxiliary total artificial heart, a persistent stagnant flow detected in the inlet port disappeared.
Design improvements to the left chamber of the auxiliary total artificial heart successfully eliminated stagnant flow at the inlet port in an in vitro model.
Answers are grounded in this paper and cite their sources.
Left chamber redesign eliminated inlet stagnation in vitro; leaves open translation to reduced thrombosis risk in auxiliary total artificial hearts.
The auxiliary total artificial heart (ATAH) is an electromechanically driven artificial heart with reduced dimensions, which is able to be implanted in the right thoracic or abdominal cavities of an average human patient without removing the natural heart or the heart neurohumoral inherent control mechanism for the arterial pressure. This device uses a brushless direct current motor and a mechanical actuator (roller screw) to move two diaphragms. The ATAH's beating frequency is regulated through the change of the left preload, based on Frank-Starling's law, assisting the native heart in obtaining adequate blood flow. The ATAH left and right stroke volumes are 38 ml and 34 ml, respectively, giving approximately 5 L/min of cardiac output at 160 bpm. Flow visualization studies were performed in critical areas on the ATAH left chamber. A closed circuit loop was used with water and glycerin (37%) at 25 degrees C. Amberlite particles (80 mesh) were illuminated by a 1 mm planar helium-neon laser light. With left mean preload fixed at 10 mm Hg and the afterload at 100 mm Hg, the heart rate varied from 60 to 200 bpm. Two porcine valves were used on the inlet and outlet ports. The flow pattern images were obtained using a color micro-camera and a video recorder. Subsequently, these images were digitized using a PC computer. A persistent stagnant flow was detected in the left chamber inlet port. After improvement on the left chamber design, this stagnant flow disappeared.
No takes yet. Share an insight, caveat, or question.
Andrade et al. (2003) studied this question. Improvement on the left chamber design of the auxiliary total artificial heart (ATAH) vs. Original left chamber design was evaluated on Flow pattern and stagnant flow in the left chamber inlet port. After improvement on the left chamber design of the auxiliary total artificial heart, a persistent stagnant flow detected in the inlet port disappeared.
Synapse has enriched 10 related papers. Consider them for comparative context: