Answers are grounded in this paper and cite their sources.
Aim. Accurate tip positioning of peripherally inserted central catheters (PICCs) is essential for safe intravenous therapy. While intracavitary ECG (IC-ECG) is a recommended method for tip confirmation, access to compatible systems may be limited in clinical practice. The aim of this case report is to describe the use of the Vygocard™ system as an alternative method for tip confirmation during the Groshong™ PICC insertion. Material and methods. We present a case of a 70-year-old oncology patient requiring removal of a totally implantable vascular access device due to infection, followed by insertion of a PICC Groshong™ for continued chemotherapy. As the Groshong™ catheter lacks a dedicated IC-ECG system, the Vygocard™ device was used with continuous saline infusion to enable electric signal conduction. Results. Real-time monitoring with Vygocard™ confirmed correct catheter tip position at the cavo-atrial junction based on P-wave morphology changes. Chest X-ray post-procedure verified accurate placement. The catheter functioned effectively and was later removed without complications. Conclusions. This case illustrates the successful off-label use of Vygocard™ for tip confirmation in a Groshong™ PICC. The approach may serve as a safe and effective alternative when dedicated systems are unavailable.
No takes yet. Share an insight, caveat, or question.
Szymczak et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: