Key result
Implantation of the paclitaxel-eluting Taxus stent for chronic total coronary occlusions resulted in a 6-month in-stent restenosis rate of 13.2% and late lumen loss of 0.13 mm.
Why the study?
Does a polymer-based paclitaxel-eluting stent improve angiographic outcomes in patients with chronic total coronary occlusions?
Observational (n=45)
Does a polymer-based paclitaxel-eluting stent improve angiographic outcomes in patients with chronic total coronary occlusions?
The use of a polymer-based paclitaxel-eluting stent appears safe and shows favorable angiographic efficacy for the treatment of chronic total coronary occlusions.
May support Taxus stent use in CTO; hypothesis-generating and requires randomized confirmation.
The purpose of this prospective nonrandomized study was to evaluate both safety and efficacy of the polymer-based paclitaxel-eluting Taxus stent for treatment of chronic total coronary occlusions. Drug-eluting stents have been proven safe and efficacious in suppressing neointimal proliferation when used in simple native de novo coronary lesions. However, there are only limited data on safety and efficacy of these stents for treatment of more difficult lesion subsets, especially chronic total occlusions. Forty-five consecutive symptomatic patients with chronic total coronary occlusions were included in this observational single-arm study. Only patients with successfully crossed occlusions were enrolled. Primary endpoints were binary restenosis and late lumen loss at 6-month angiographic follow-up. Secondary endpoints were MACE at 30-day and 6-month follow-up. The 30-day MACE rate was 0%. At 6 months, the cumulative MACE-free survival was 84.4%. There were no deaths, myocardial infarctions, or stent thrombosis up to 6 months poststenting. At angiographic follow-up, the in-stent restenosis rate was 13.2% (5/38). There was a total of five in-stent restenoses, with three focal and two diffuse restenosis patterns. The in-segment late lumen loss was 0.13 +/- 0.58 mm. The stent edge analyses revealed a late loss of 0.21 +/- 0.66 proximally and a negative late loss of -0.10 +/- 0.67 at the distal edge segment. This study shows safety and indicators of efficacy of implantation of the paclitaxel-eluting Taxus stent for treatment of chronic total coronary occlusions. The quantitative angiographic analysis revealed beneficial results with respect to the efficacy surrogates binary restenosis rate and late loss.
No takes yet. Share an insight, caveat, or question.
Buellesfeld et al. (2005) conducted an observational in chronic total coronary occlusions (n=45). Polymer-based paclitaxel-eluting Taxus stent was evaluated on binary restenosis and late lumen loss at 6-month angiographic follow-up. Implantation of the paclitaxel-eluting Taxus stent for chronic total coronary occlusions resulted in a 6-month in-stent restenosis rate of 13.2% and late lumen loss of 0.13 mm.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: