Key result
The AMISA device with a ready-to-use collagen fleece demonstrated significantly higher adhesive stability for selective leak closure of porcine pleura in vitro compared with liquid gluing methods.
Answers are grounded in this paper and cite their sources.
The AMISA device provides a novel method for applying tissue glue via collagen fleece in minimally invasive surgery, demonstrating superior adhesive stability in preliminary in vitro models.
SummaryAdjustable minimally-invasive surgery applicator (AMISA) is an innovative, reusable instrument developed for tissue gluing, e.g. for sealing large defects on parenchymal organs. The device AMISA is loaded with TachoCombr`, a commercially-available, ready-to-use dry collagen fleece, coated with tissue glue. After loading the adjustable applicator head with the 5 x10 cm, self-adhesive collagen fleece, the AMISA can be introduced into the body through a conventional 10/12 mm trocar. The fleece roll is positioned over the tissue defect and can be rolled-out (‘carpeting’) utilising the rotary positioning of the applicator head. Preliminary technological studies of selective leak closure (SLC) have demonstrated the efficiency of fleece-bound tissue gluing with a ready-to-use collagen fleece, coated with tissue glue. In vitro studies of SLC of porcine pleura have demonstrated significantly higher adhesive stability compared with liquid gluing methods employing various fleece materials. AMISA can be used in both thoracoscopy and laparoscopy.The main indications are large-area leakages on parenchymal organs of differing aetiology that cannot be controlled speedily and efficiently with conventional minimally-invasive technology (mechanical, thermal, piezoelectric). The AMISA system expands the range of applications of minimally-invasive surgery and simplifies complication management (bleeding, rupture, leakage, secretion). Other advantages of the system are decrease in drainage time and reduction in treatment time.
No takes yet. Share an insight, caveat, or question.
Carbon et al. (1999) studied Large-area leakages on parenchymal organs. Adjustable minimally-invasive surgery applicator (AMISA) vs. Liquid gluing methods was evaluated on Adhesive stability. The AMISA device with a ready-to-use collagen fleece demonstrated significantly higher adhesive stability for selective leak closure of porcine pleura in vitro compared with liquid gluing methods.
Synapse has enriched 10 related papers. Consider them for comparative context: