Key result
The ECG method for positioning right atrial catheters achieved similar accuracy to fluoroscopy (average 1.9 vs 1.1 cm) while reducing operating room time by 3.0 minutes (p < 0.04).
Why the study?
Does an ECG technique maintain accuracy and reduce time/cost compared to fluoroscopy for positioning right atrial catheters?
Observational (n=1,822)
Does an ECG technique maintain accuracy and reduce time/cost compared to fluoroscopy for positioning right atrial catheters?
Answers are grounded in this paper and cite their sources.
The ECG method is a safe, accurate, and cost-effective alternative to fluoroscopy for positioning right atrial catheters.
OBJECTIVE: This study evaluated the accuracy of placing right atrial catheters using an electrocardiographic (ECG) technique. SUMMARY BACKGROUND DATA: Placement of right atrial catheters for vascular access is a common operative procedure. Accurate placement is essential for proper function. Previous placement techniques have used fluoroscopy, which is both time consuming and hazardous. METHODS: The accuracy of placement of 1236 right atrial catheters using an ECG technique was compared to placement of 586 catheters using fluoroscopy between March 1991 and November 1995. In the ECG technique, the catheter was flushed with sodium bicarbonate. A sterile left-leg ECG lead was attached to the catheter with the other ECG leads applied normally. On advancing the catheter through the superior vena cava, the P-wave amplitude (lead II) increased in negative deflection until greater than the QRS complex. Passing the sinoatrial node, the P-wave developed an initial positive then negative deflection. The catheter was positioned so the P-wave was biphasic, representing a position midway between the sinoatrial and atrioventricular nodes. For the fluoroscopic technique, catheters were positioned under direct observation just within the atrium estimated from cardiac contour. Use of contrast was optional if atrial anatomy was unclear. RESULTS: Postoperative portable chest x-rays showed the ECG method to position the catheter tip within the right atrium just as accurately (average, 1.9 +/- 1.3 cm) as with the use of fluoroscopy (average, 1.1 +/- 1.6 cm). The ECG method eliminated an average of 20 seconds of radiation exposure, an average of 3.0 minutes operating room time (p < 0.04), avoided all risks of contrast dye, and saved $279.10 per case. CONCLUSIONS: The ECG method is a satisfactory alternative to that of fluoroscopy for placement of long-term central venous catheters into the right atrium.
No takes yet. Share an insight, caveat, or question.
Watters et al. (1997) conducted an observational in Vascular access requiring right atrial catheter placement (n=1,822). Electrocardiographic (ECG) technique vs. Fluoroscopy was evaluated on Accuracy of catheter tip position within the right atrium (cm). The ECG method for positioning right atrial catheters achieved similar accuracy to fluoroscopy (average 1.9 vs 1.1 cm) while reducing operating room time by 3.0 minutes (p < 0.04).
Synapse has enriched 10 related papers. Consider them for comparative context: