Answers are grounded in this paper and cite their sources.
The diffusion of O2 and CO2 through the walls of two types of plastic syringes was measured. When the syringes were filled with 16% O2, 5% CO2, and the balance N2, the average outward diffusion of CO2 was 0.076% per hour and the average inward diffusion of O2 was 0.06% per hour in room air at room temperature. When the syringes were surrounded by room air cooled in ice the average loss of CO2 was 0.137% per hour and the average increase in O2 content 0.10% per hour. From the measurements we conclude that diffusion of gases through the walls of disposable syringes cannot be responsible for the low [Formula: see text] values of "normal" arterial blood reported by clinical laboratories.
No takes yet. Share an insight, caveat, or question.
Rosenberg et al. (1973) studied this question.