A gradient reversed-phase column high-performance liquid chromatographic method was developed for the detection and quantification of norfloxacin and its major impurities in norfloxacin-containing pharmaceuticals. Chromatographic separations were performed under the following experimental conditions: column, Zorbax SB RP-18 (5 microm, 250 x 4.6 mm); injection volume, 20 microL; mobile phase, 0.05 M NaH2PO4 (pH 2.5)-acetonitrile (87 + 13) for 16 min and (58 + 42) for 9 min (stepwise gradient); and flow rate, 1.3 mL/min. All analyses were performed at 25 degrees C, and the eluate was monitored at 275 nm using a diode array detector. Linearity (correlation coefficient = 0.999), recovery (99.3-101.8%), relative standard deviation (0.2-0.7%), and quantitation limit (0.12-0.47 microg/mL) were evaluated and found to be satisfactory. The method is simple, rapid, and convenient for purity control of norfloxacin in both raw materials and dosage forms.
No takes yet. Share an insight, caveat, or question.
Mielji et al. (2008) studied this question.