Answers are grounded in this paper and cite their sources.
A new multiturn tandem time-of-flight (TOF) mass spectrometer "MULTUM-TOF/TOF" has been designed and constructed. It consists of a matrix-assisted laser desorption/ionization ion source, a multiturn TOF mass spectrometer, a collision cell, and a quadratic-field ion mirror. The multiturn TOF mass spectrometer can overcome the problem of precursor ion selection in TOF, due to insufficient time separation between two adjacent TOF peaks, by increasing the number of cycles. As a result, the total TOF increases with the increase in resolving power. The quadratic-field ion mirror allows temporal focusing for fragment ions with different kinetic energies. Product ion spectra from monoisotopically selected precursor ions of angiotensin I, substance P, and bradykinin have been obtained. The fragment ions observed are mainly the result of high-energy collision induced dissociation.
No takes yet. Share an insight, caveat, or question.
Toyoda et al. (2007) studied this question.