Answers are grounded in this paper and cite their sources.
The precise quantification of oxygen concentration in titanium is crucial for various high-performance applications. In this study, we developed reference samples to calibrate high-resolution micro laser-induced breakdown spectroscopy (HR-µLIBS) for oxygen analysis in titanium. Two fabrication methods were employed: (i) oxidation of titanium powder at high temperature in a controlled atmosphere with oxygen, followed by spark plasma sintering (SPS), and (ii) blending titanium powder with TiO 2 powder before sintering by SPS. Homogenized samples were then used to establish a calibration line correlating HR-µLIBS intensity with known oxygen concentrations. To evaluate the HR-µLIBS method, a titanium sample oxidized 500 h at 650°C in Ar-20%O 2 was analyzed using both HR-µLIBS and electron probe microanalysis (EPMA). The results demonstrate a strong agreement between the two methods, with HR-µLIBS offering superior speed and improved accuracy at low oxygen concentrations (< 3 at. %).
No takes yet. Share an insight, caveat, or question.
Gautier et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: