Key result
Computed tomography of the chest successfully diagnosed a subacute right ventricular pacemaker lead perforation presenting as left hemothorax and shock, which had been missed by chest X-ray and echocardiogram.
Why the study?
Pacemaker lead-induced heart perforation is a rare but life-threatening complication of pacemaker implantation, and timely diagnosis remains a challenge for clinicians.
Case Report (n=1)
No
Normal pacemaker function on device interrogation does not rule out lead perforation; advanced imaging like CT scan or POCUS should be utilized when perforation is suspected but missed by standard X-ray and echocardiogram.
Normal interrogation does not exclude perforation; this case leaves open POCUS/CT utility when standard tests are negative.
BACKGROUND: Pacemaker lead-induced heart perforation is a rare but life-threatening complication of pacemaker implantation, and timely diagnosis remains a challenge for clinicians. Here, we report a case of pacemaker lead-induced cardiac perforation rapidly diagnosed by a "bow-and-arrow" sign on point-of-care ultrasound (POCUS). CASE SUMMARY: A 74-year-old Chinese woman who had undergone permanent pacemaker implantation 26 d before suddenly developed severe dyspnea, chest pain, and hypotension. The patient had received emergency laparotomy for an incarcerated groin hernia and was transferred to the intensive care unit 6 d before. Computed tomography was not available due to unstable hemodynamic status, so POCUS was performed at the bedside and revealed severe pericardial effusion and cardiac tamponade. Subsequent pericardiocentesis yielded a large volume of bloody pericardial fluid. Further POCUS by an ultrasonographist revealed a unique "bow-and-arrow" sign indicating right ventricular (RV) apex perforation by the pacemaker lead, which facilitated the rapid diagnosis of lead perforation. Given the persistent drainage of pericardial bleeding, urgent off-pump open chest surgery was performed to repair the perforation. However, the patient died of shock and multiple organ dysfunction syndrome within 24 h post-surgery. In addition, we also performed a literature review on the sonographic features of RV apex perforation by lead. CONCLUSION: POCUS enables the early diagnosis of pacemaker lead perforation at the bedside. A step-wise ultrasonographic approach and the "bow-and-arrow" sign on POCUS are helpful for rapid diagnosis of lead perforation.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2023) conducted a case report in Pacemaker lead-induced heart perforation (n=1). Computed tomography (CT) of the chest vs. Chest X-ray and echocardiogram was evaluated on Diagnosis of pacemaker lead perforation. Computed tomography of the chest successfully diagnosed a subacute right ventricular pacemaker lead perforation presenting as left hemothorax and shock, which had been missed by chest X-ray and echocardiogram.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: