Traditional full-field optical coherence tomography (FF-OCT) systems face limitations in imaging efficiency and data redundancy, particularly when applied to complex biological samples. To address these issues, we developed an automated FF-OCT system that uses a variable resolution z-scanning strategy to dynamically adjust the voxel resolution of acquired images according to sample structures. The system was validated using 3D HepaRG cell cultures embedded in micro-hydrogels. To optimize imaging efficiency and minimize data redundancy, an in-air voxel resolution of 0.7×0.7×5µm 3 was applied to the region of interest (ROI) for detailed sample analysis, while a lower resolution of 1.4×1.4×10µm 3 was used in non-ROI areas. Compared to traditional equidistant acquisition methods, the variable resolution strategy reduced imaging time by over 20% and data storage requirements by more than 35%, with deviations in morphological parameters, including volume and surface area, kept below 1%. Robustness tests across multiple cell culture batches confirmed the system’s reliability in accurately capturing complex biological structures. This study demonstrates a significant advancement in FF-OCT technology, providing a practical, high-efficiency, and high-precision solution for non-invasive imaging of complex biological samples.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: