Key result
Direct-current shocks delivering 194±11 joules terminated 98% of ventricular fibrillation episodes in patients weighing 91 to 225 kg, indicating that higher energy shocks are unnecessary.
Why the study?
Do standard energy direct-current shocks effectively terminate ventricular fibrillation in heavy patients?
Observational (n=12)
Do standard energy direct-current shocks effectively terminate ventricular fibrillation in heavy patients?
Answers are grounded in this paper and cite their sources.
Standard energy defibrillation (~194 joules delivered) is highly effective (98%) for terminating ventricular fibrillation even in heavy patients, challenging the need for high-energy (400-800 joules) defibrillators.
In a prospective survey employing conventional devices that stored 400 joules or less, direct-current shocks that delivered 194±11 joules (1.8 joules/kg) to the chest wall terminated 45 of 46 episodes of ventricular fibrillation in 11 of 12 patients weighing 91 to 225 kg. Patients with coronary disease defibrillated more easily than patients without coronary disease. Weight did not determine outcome. The 98% defibrillation efficiency occurred at only half the energy predicted for 50% efficiency from retrospective data. Human defibrillation not only occurred at one third the untested level recommended by proponents of high-energy defibrillation, but also at one fourth the energy, defibrillating 69% of comparably heavy animals. Thus, the costly defibrillators delivering 400 to 800 joules now sold by 11 of 14 American manufacturers are superfluous, untested, potentially lethal devices with which to attempt ventricular defibrillation. (JAMA242:1380-1384, 1979).
No takes yet. Share an insight, caveat, or question.
J. A. Gascho (1979) conducted an observational in Ventricular fibrillation (n=12). Direct-current shocks delivering 194±11 joules was evaluated on Termination of ventricular fibrillation episodes. Direct-current shocks delivering 194±11 joules terminated 98% of ventricular fibrillation episodes in patients weighing 91 to 225 kg, indicating that higher energy shocks are unnecessary.