This is the authors' abstract. We don't add key points for this paper.
Optomechanical accelerometers offer in situ traceability to the international system of units through laser interferometry, providing an alternative to a calibration chain using instrumented shakers. Here, we examine the ‘self-calibrating’ property of a prototype optomechanical accelerometer for use as a seismic reference. We report the optomechanically derived sensitivity of the accelerometer and compare this in situ calibrated output to input accelerations from an instrumented shaker, finding agreement to be within ±1%. The comparison spanned frequencies between 3 Hz and 30 Hz, and for sinusoidal accelerations with amplitudes ranging from 0.01 m s−2 to 0.6 m s−2. These results are evidence that optomechanically derived sensitivity calibration can be equivalent to established international methods for primary calibration.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Pratt et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: