Answers are grounded in this paper and cite their sources.
An ion-selective electrode (ISE) was developed for the rapid determination of pseudoephedrine hydrochloride (PSEHCl) in pharmaceutical preparations. The electrode incorporates a PVC membrane with a pseudoephedrine–phosphotungstate ion pair complex. The influences of membrane composition, temperature, pH of the test solution, and the interfering ions on the electrode performance were investigated. The sensor exhibits a Nernstian response for pseudoephedrine hydrochloride ions over a relatively wide concentration range (1.0 × 10−1 to 1.0 × 10−5 mol L−1) with a slope of 56.2 ± 0.5 mV per decade at 25°C. It can be used in the pH range 4.0–10.5. The isothermal temperature coefficient of this electrode amounted to 0.0009 V/°C. The membrane sensor was successfully applied to determination of PSEHCl in its tablets and syrup.
No takes yet. Share an insight, caveat, or question.
Giahi et al. (2009) studied this question.