Key result
DES account for ~89% of MAUDE adverse event reports, showing rising malfunction rates over time.
Why the study?
Although extensive research exists on perioperative outcomes of bare-metal stents (BMS), drug-eluting stents (DES), and bioresorbable drug-eluting stents (BVS), a knowledge gap remained in the longitudinal monitoring of patient outcomes due to device-related causes.
What are the longitudinal trends in device-related adverse outcomes (deaths, injuries, malfunctions) and misreporting rates for drug-eluting stents, bare-metal stents, and bioresorbable drug-eluting stents in the FDA MAUDE database?
Observational (n=68,618)
Yes
What are the longitudinal trends in device-related adverse outcomes (deaths, injuries, malfunctions) and misreporting rates for drug-eluting stents, bare-metal stents, and bioresorbable drug-eluting stents in the FDA MAUDE database?
Longitudinal FDA MAUDE analysis reveals a 4-fold greater increase in reported DES malfunctions relative to the decrease in BMS malfunctions, an inverse shift between DES injuries and malfunctions, and a 7% death misreporting rate, underscoring the need for standardized device outcome reporting.
Answers are grounded in this paper and cite their sources.
Rising DES malfunction reports in MAUDE data should not change stent selection; leaves open reporting bias versus true incidence.
The growing prevalence of coronary artery diseases in the US corresponds to the increasing use of minimally invasive techniques that require coronary stents. Although extensive research is available on the perioperative outcomes of the 3 stent options - bare-metal stents (BMS), drug-eluting stents (DES), and bioresorbable drug-eluting stents (BVS), a knowledge gap exists in the longitudinal monitoring of patient outcomes due to device-related causes. Therefore, our study examines the device-related patient outcome and the relative performance for BMS, DES, and BVS. Data on 3 device outcomes (deaths, injuries, and malfunction) for each stent type was obtained from the January 2011 to February 2020 Manufacturer and User Facility Device Experience (MAUDE) database. Statistical visualizations and analysis were used to identify trends and significant differences between groups. Of a total of 68,618 adverse event reports, DES, BMS, and BVS each accounted for 88.5 %, 10.2 %, and 1.25 % of the cases, respectively. Device malfunctions were the most reported event (47.2 %), followed by injuries (44.1 %) and deaths (8.66 %). Over time, BMS malfunction rates showed a steady decrease (R = -0.87), while DES malfunction rates increased significantly (R = 0.79). An inversely proportional relationship between DES injuries and malfunctions was observed. The increase in DES malfunctions was 4 times greater than the decrease in BMS malfunctions. Approximately 7 % of reported adverse events were classified as misreported, with most involving DES. These results suggest 2 plausible interpretations: 1) reporting categorization for devices shifted from injuries to malfunction, and 2) stents choice is transitioning from BMS to DES. Our findings also highlight the need to improve reporting accuracy for MAUDE database data.
No takes yet. Share an insight, caveat, or question.
Gao et al. (2024) conducted an observational in Coronary artery disease treated with coronary stents (n=68,618). Drug-eluting stents (DES) and bioresorbable drug-eluting stents (BVS) vs. Bare-metal stents (BMS) was evaluated on Device-related adverse event outcomes (malfunctions, injuries, and deaths). Drug-eluting stents accounted for 88.5% of 68,618 MAUDE stent reports, showing a significant increase in malfunction rates over time vs a steady decrease for bare-metal stents.
Synapse has enriched 10 related papers. Consider them for comparative context: