In a cohort of unselected patients, implantable cardioverter-defibrillators equipped with SmartShock technology accurately categorized ventricular tachyarrhythmias with a positive predictive value of 99.5%, and antitachycardia pacing successfully treated 99.9% of episodes.
Why the study?
Little is known about ICD performance using enhanced detection algorithms in unselected, non-trial patients.
Observational (n=196)
Yes
In a real-world cohort, ICDs equipped with enhanced detection algorithms accurately identified ventricular tachyarrhythmias and effectively treated them with antitachycardia pacing, resulting in a low burden of inappropriate shocks.
BACKGROUND: Little is known about the ICD performance using enhanced detection algorithms in unselected, non-trial patients. Performance of recent generation ICD equipped with SmartShock™ technology (SST) for detection and conversion of ventricular tachyarrhythmias (VTA) was investigated. METHODS: 4P was a prospective, multicenter, observational study conducted in 10 Swiss implanting centers. Patients with a Class I indication according to international guidelines were included and received an ICD with SST. ICD discrimination capability was assessed by evaluating SST performance; therapy efficacy was assessed by rate of VTA conversions by ATP and by rescue shocks. RESULTS: Overall, 196 patients were included in the analysis with a mean duration of follow-up of 27.7 months (452 patient-years of observation). Patient-specific rather than recommended programming was preferred. Device-detected episodes were frequent (5147 episodes in 146 patients, 74.5%). In 44 patients (22.4%), 1274 episodes were categorized as VTA; only 215 episodes were symptomatic. ATP was the first-line therapy and highly effective (99.9% success rate at the episode level, 100.0% at the patient level). Rescue shocks were rare (66 episodes in 28 patients); 7 shocks in 5 patients (2.6%) were inappropriate. Death and hospitalization rates were low. CONCLUSIONS: In a cohort of non-trial, unselected ICD patients, VTA episodes were frequent. The 4P results confirm the robustness of VTA detection by SST and the effectiveness of ATP treatment, hence limiting overall ICD shock burden.
No takes yet. Share an insight, caveat, or question.
Regoli et al. (2019) conducted an observational in Class I indication for primary or secondary prevention of sudden cardiac death (n=196). Implantable cardioverter-defibrillator with SmartShock technology was evaluated on Antitachycardia pacing (ATP) success rate at the episode level. In a cohort of unselected patients, implantable cardioverter-defibrillators equipped with SmartShock technology accurately categorized ventricular tachyarrhythmias with a positive predictive value of 99.5%, and antitachycardia pacing successfully treated 99.9% of episodes.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: