A combination of electrospray ionization, high-field asymmetric waveform ion mobility spectrometry, and mass spectrometry (ESI-FAIMS/MS) was used to analyze standard solutions of microcystins-LR, -RR, and -YR. The ability of FAIMS to separate ions in the gas phase reduced the amount of background in the mass spectrum without compromising the absolute signal for these microcystins. This reduction in background resulted in a ten-fold improvement in the signal-to-background ratio over conventional ESI-MS. Detection limits, using direct infusion, were determined to be 4, 2, and 1 nM for microcystins-LR, -RR, and -YR, respectively.