Key result
In a 3D printed phantom model, CT scanning with high pitch and 100 kVp effectively reduced radiation dose while maintaining image quality for post-TEVAR surveillance.
Why the study?
Surveillance CT scans in post-TEVAR patients are associated with high radiation doses, carrying a potential risk of radiation-induced malignancy.
What are the optimal CT scanning parameters (kVp and pitch) to minimize radiation dose while maintaining image quality in post-TEVAR patients?
What are the optimal CT scanning parameters (kVp and pitch) to minimize radiation dose while maintaining image quality in post-TEVAR patients?
Optimizing CT angiography protocols using 100 kVp and high pitch can reduce radiation exposure without compromising image quality in post-TEVAR surveillance.
Patient-specific phantoms may guide lower-dose CT protocols post-TEVAR; leaves open clinical validation in TBAD surveillance.
Thoracic endovascular aortic repair (TEVAR) is a life-saving therapy for type B aortic dissection (TBAD). However, surveillance computed tomography (CT) scans in post-TEVAR patients are associated with high radiation dose, thus resulting in potential risk of radiation-induced malignancy. In this study, we developed a patient-specific three-dimensional (3D) printed phantom with stent grafts in situ, then scanned the phantom with different CT protocols to determine the optimal scanning parameters for post-treatment patients. The CT scans were conducted with different kVp and pitch values (80, 100, 120 kVp and pitch of 1.2, 1.5, 2.0, 2.5), resulting in a total of 12 datasets. Signal-to-noise ratio (SNR) was measured to determine and compare the image quality between different datasets. Results showed no significant differences in SNR between different kVp when the pitch value was 1.2. At low pitch values, a decrease in kVp from 120 to 80 led to a significant effective dose reduction by more than 20%. SNR decreased by 30% when pitch was increased from 1.2 to 2.5 at 80 kVp, and 20% at 120 kVp. In contrast, there was only a 3.9% decrease in SNR when kVp was reduced from 120 to 80 at pitch 1.2, and 15.9% at pitch 2.5. High pitch with 100 kVp can effectively reduce the dose while maintaining image quality.
No takes yet. Share an insight, caveat, or question.
Wu et al. (2021) studied Type B aortic dissection. CT scanning with different kVp and pitch values was evaluated on Signal-to-noise ratio (SNR) and effective dose. In a 3D printed phantom model, CT scanning with high pitch and 100 kVp effectively reduced radiation dose while maintaining image quality for post-TEVAR surveillance.