Key result
Short-pulse mode microwave ablation yielded no significant differences (<5%) in coagulation zone diameter, volume, or sphericity compared to continuous mode in computer simulations and ex vivo models.
Why the study?
The differences between continuous and short-pulse mode microwave ablation were not clearly established.
Does short-pulse mode microwave ablation improve coagulation zone size and sphericity compared to continuous mode in computational and ex vivo models?
Does short-pulse mode microwave ablation improve coagulation zone size and sphericity compared to continuous mode in computational and ex vivo models?
Short-pulse mode microwave ablation does not significantly enlarge coagulation zones or improve sphericity compared to continuous mode in computational and ex vivo models.
Answers are grounded in this paper and cite their sources.
Short-pulse MWA may alter coagulation zones versus continuous mode in models; leaves open optimal parameters for clinical use.
Purpose To study the differences between continuous and short-pulse mode microwave ablation (MWA).Methods We built a computational model for MWA including a 200 mm long and 14 G antenna from Amica-Gen and solved an electromagnetic-thermal coupled problem using COMSOL Multiphysics. We compared the coagulation zone (CZ) sizes created with pulsed and continuous modes under ex vivo and in vivo conditions. The model was used to compare long vs. short pulses, and 1000 W high-powered short pulses. Ex vivo experiments were conducted to validate the model.Results The computational models predicted the axial diameter of the CZ with an error of 2–3% and overestimated the transverse diameter by 9–11%. For short pulses, the ex vivo computer modeling results showed a trend toward larger CZ when duty cycles decreases. In general, short pulsed mode yielded higher CZ diameters and volumes than continuous mode, but the differences were not significant (<5%), as in terms of CZ sphericity. The same trends were observed in the simulations mimicking in vivo conditions. Both CZ diameter and sphericity were similar with short and long pulses. Short 1000 W pulses produced smaller sphericity and similar CZ sizes under in vivo and ex vivo conditions.Conclusions The characteristics of the CZ created by continuous and pulsed MWA show no significant differences from ex vivo experiments and computer simulations. The proposed idea of enlarging coagulation zones and improving their sphericity in pulsed mode was not evident in this study.
No takes yet. Share an insight, caveat, or question.
Radosevic et al. (2021) studied this question. Short-pulse mode microwave ablation vs. Continuous mode microwave ablation was evaluated on Coagulation zone (CZ) size and sphericity. Short-pulse mode microwave ablation yielded no significant differences (<5%) in coagulation zone diameter, volume, or sphericity compared to continuous mode in computer simulations and ex vivo models.
Synapse has enriched 10 related papers. Consider them for comparative context: