Key result
MR-compatible microwave ablation shows no difference in ablation diameter or volume vs standard systems.
Why the study?
Evaluation of a newly developed MR-compatible microwave ablation system was needed to assess ablation performance compared to a standard microwave ablation system.
Does an MR-compatible microwave ablation system provide comparable ablation performance to a standard microwave ablation system in an ex vivo bovine liver model?
Does an MR-compatible microwave ablation system provide comparable ablation performance to a standard microwave ablation system in an ex vivo bovine liver model?
The newly developed MR-compatible microwave ablation system performs comparably to the standard system in an ex vivo bovine liver model, suggesting its viability for MR-guided interventions.
Answers are grounded in this paper and cite their sources.
MR-compatible MWA performs comparably to standard systems ex vivo; leaves open clinical adoption pending in vivo and human validation.
OBJECTIVE: Evaluation of a newly developed MR-compatible microwave ablation system with focus on ablation performance and comparison with a corresponding standard microwave ablation system. MATERIALS AND METHODS: A total of 52 ablations were performed with a non-cooled microwave ablation system in an ex vivo bovine liver model using the following settings: [A] 16G-standard antenna, 2 cm active tip, 2.4 m cable; [B] MR-compatible 16G-antenna, 2 cm active tip, 2.4 m cable; [C] MR-compatible 16G-antenna, 2 cm active tip, extended 6 m cable; and [D] MR-compatible 16G-antenna, 4 cm active tip, extended 6 m cable. Ablation durations were 3, 5 and 10 min, and additionally 15 min for [D]. Ablations zones were measured for short-axis diameter (SA) and long-axis diameter (LA). Settings [A]-[C] were compared regarding SA, volume (V) and generator energy output (E) with analysis of variance and Tukey-Kramer post hoc test. Ablation performance of the MR-compatible settings [C] and [D] were compared regarding SA, V, E and sphericity index (SA/LA) with unpaired t-test. p < 0.05 was considered as statistically significant. RESULTS: No significant differences were found between [A], [B] and [C] regarding SA and V (10 min; SA[A] = 25.8 ± 2.4 mm, SA[B] = 25.3 ± 1.9 mm, SA[C] = 25.0 ± 2.0 mm, p = 0.88; V[A] = 17.8 ± 4.4 cm³, V[B] = 16.6 ± 3.0 cm³, V[C] = 17.8 ± 2.7 cm³, p = 0.85); however, the highest energy output was measured for setting [C] (10 min; [A]: 9.9 ± 0.5 kJ, [B]: 10.1 ± 0.5 kJ, [C]: 13.1 ± 0.3 kJ, p < 0.001). SA, V and E were significantly larger with setting [D] than [C] with 10 min ablations (SA[C] = 25.0 ± 2.0 mm, SA[D] = 34.0 ± 2.9 mm, p = 0.003; V[C] = 17.8 ± 2.7 cm³, V[D] = 39.4 ± 7.5 cm³, p = 0.007; E[C] = 13.1 ± 0.3 kJ, E[D] = 16.7 ± 0.8 kJ, p = 0.002) without significant difference in sphericity index (SA/LA[C] = 0.46 ± 0.02, SA/LA[D] = 0.52 ± 0.04, p = 0.08). CONCLUSION: The tested MR-compatible system can be used without loss of ablation performance compared to the standard system.
No takes yet. Share an insight, caveat, or question.
Hoffmann et al. (2017) studied Ex vivo bovine liver model (n=52). MR-compatible microwave ablation system vs. Standard microwave ablation system was evaluated on Short-axis diameter (SA) and volume (V) of ablation zones (p=0.88). The MR-compatible microwave ablation system demonstrated no significant differences in ablation short-axis diameter or volume compared to the standard system.