Answers are grounded in this paper and cite their sources.
BACKGROUND: The objective of this study was to determine the peak contact pressure with varying degrees of articular cartilage step-off in a transtectal acetabular fracture model. METHODS: Five fresh frozen cadaveric hip joints were potted in a custom loading fixture. The five specimens were then tested at loads of 445 N (newton) (100 lb) and 1,335 N (300 lb) intact and after a transverse osteotomy at step-off levels from 0 to 5 mm in 1-mm increments. RESULTS: Articular cartilage step-off of greater than 1 mm led to significantly increased contact stress at the loaded acetabular articular surface. Mean peak pressure measured at 1,335 N of loading in all intact specimens before the osteotomy was approximately 10 MPa. Peak pressure after a transverse acetabular fracture did not change when the fracture was perfectly reduced. At 1 mm of step-off, the peak pressure increased by approximately 20% but was not statistically significant. With step-off of > 2 mm or greater, the peak pressure increase was approximately 50% and was statistically significant. CONCLUSION: On the basis of our study, transverse acetabular fractures with greater than 1 mm of displacement can lead to significant increase in peak pressure at the articular surface.
No takes yet. Share an insight, caveat, or question.
Malkani et al. (2001) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: