A flow injection analysis (FIA) procedure for the speciation of Cr(III) and Cr(VI) using the 1, 5-diphenylcarbazide (DPC) method is presented. As Cr(III) does not interfere in the Cr(VI)-DPC reaction, both Cr(VI) and total chromium [after the on-line oxidation of Cr(III) by Ce(V)] are sequentially determined. Cr(III) is obtained by difference. Under the experimental conditions described, the calibration graphs are linear up to 2 µg ml–1 of Cr(VI) and 4 µg ml–1 of Cr(III). The detection limits found were 18 ng ml–1 for Cr(VI) and 55 ng ml–1 for Cr(III), at a signal to noise ratio of 3. The common interfering elements in the Cr(VI)-DPC reaction were investigated under dynamic FIA conditions. The FIA method was also compared with the conventional spectrophotometric procedure.
No takes yet. Share an insight, caveat, or question.
Rocha et al. (1985) studied this question.