Key result
Local saline perfusion during radiofrequency ablation increased the ablation zone longitudinal diameter, and estimated volume increased only when double passage was performed.
Why the study?
Radiofrequency ablation induces variable ablation zones depending on technique and organ impedance, with a rationale for its use in canine liver tumours.
Does local saline perfusion and double passage affect the size and geometry of radiofrequency ablation zones in an ex vivo swine liver model?
Does local saline perfusion and double passage affect the size and geometry of radiofrequency ablation zones in an ex vivo swine liver model?
In an ex vivo swine liver model, local saline perfusion during radiofrequency ablation increased the longitudinal diameter of the ablation zone, with volume increases seen only after double passage.
Ex vivo swine data on RFA zone reproducibility and saline effects inform technique choice in canine liver tumors; leaves open need for in vivo trials before practice change.
BACKGROUND: Radiofrequency Ablation (RFA) is rarely performed in veterinary medicine. A rationale exists for its use in selected cases of canine liver tumours. RFA induces ablation zones of variable size and geometry depending on the technique used and on the impedance of the targeted organ. OBJECTIVES: (a) to describe the geometry and reproducibility of the ablation zones produced by three commercially available systems from a single company, using isolated swine liver parenchyma as a model for future veterinary applications in vivo; (b) to study the effects of local saline perfusion into the ablated parenchyma through the electrode tip and of single versus double passage of the electrode on size, geometry and reproducibility of the ablation zones produced. METHODS: Size, and geometry of ablation zones reproduced in six livers with one cooled and perfused (saline) and two cooled and non-perfused systems, after single or double passage (n = 6/condition), were assessed macroscopically on digitalized images by a blinded operator. Longitudinal and transverse diameters, equivalent diameter, estimated volume and roundness index were measured. Reproducibility was assessed as coefficient of variation. RESULTS AND CONCLUSIONS: Ablation zone reproducibility was higher when expressed in terms of ablation zone diameters than estimated volume. Local saline perfusion of the parenchyma through the electrode tip during RFA increased the ablation zone longitudinal diameter. Ablation zone estimated volume increased with saline perfusion only when double passage was performed. These data may provide useful information for those clinicians who intend to include RFA as an additive tool in veterinary interventional radiology.
No takes yet. Share an insight, caveat, or question.
Bullone et al. (2020) studied Ex vivo swine liver model (n=6). Radiofrequency ablation with local saline perfusion and single vs double passage vs. Cooled and non-perfused radiofrequency ablation systems was evaluated on Size, geometry, and reproducibility of ablation zones (diameters, estimated volume, roundness index). Local saline perfusion during radiofrequency ablation increased the ablation zone longitudinal diameter, and estimated volume increased only when double passage was performed.