Key result
Numerical simulation of two composite aortic prostheses showed that vortical structures do not influence bileaflet valve kinematics, but major differences exist in stress status near the suture line.
Numerical simulations of composite aortic prostheses reveal that while graft type does not affect valve kinematics or coronary flow, it significantly impacts stress at the coronary suture lines.
Suture-line stress differences may affect durability; hypothesis-generating for composite graft selection in aortic surgery.
The in vivo evaluation of prosthetic device performance is often difficult, if not impossible. In particular, in order to deal with potential problems such as thrombosis, haemolysis, etc., which could arise when a patient undergoes heart valve replacement, a thorough understanding of the blood flow dynamics inside the devices interacting with natural or composite tissues is required. Numerical simulation, combining both computational fluid and structure dynamics, could provide detailed information on such complex problems. In this work, a numerical investigation of the mechanics of two composite aortic prostheses during a cardiac cycle is presented. The numerical tool presented is able to reproduce accurately the flow and structure dynamics of the prostheses. The analysis shows that the vortical structures forming inside the two different grafts do not influence the kinematics of a bileaflet valve or the main coronary flow, whereas major differences are present for the stress status near the suture line of the coronaries to the prostheses. The results are in agreement with in vitro and in vivo observations found in literature.
No takes yet. Share an insight, caveat, or question.
Tullio et al. (2011) studied Heart valve replacement. Numerical simulation of two composite aortic prostheses was evaluated on Flow and structure dynamics, including vortical structures and stress status. Numerical simulation of two composite aortic prostheses showed that vortical structures do not influence bileaflet valve kinematics, but major differences exist in stress status near the suture line.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: