Key result
Daily implant-based multiparametric telemonitoring significantly reduced the number of in-hospital device controls by 44% (p < 0.01) in heart failure patients with a ventricular assist device.
Why the study?
Telemonitoring of heart failure patients is becoming increasingly important, but its benefit in end-stage heart failure patients with a ventricular assist device required evaluation.
Does daily implant-based multiparametric telemonitoring reduce in-hospital device controls and detect arrhythmias early in end-stage HF patients with a VAD?
Cohort (n=26)
Does daily implant-based multiparametric telemonitoring reduce in-hospital device controls and detect arrhythmias early in end-stage HF patients with a VAD?
Daily telemonitoring in VAD patients with an ICD or CRT-D safely reduces in-hospital device checks and enables early detection of arrhythmias and lead complications.
The telemonitoring of heart failure (HF) patients is becoming increasingly important. This study aimed to evaluate the benefit of telemonitoring in end-stage HF patients with a ventricular-assistance device (VAD). A total of 26 HF-patients (66 ± 11 years, 88% male) on VAD therapy with an implantable cardioverter-defibrillator (ICD) or a cardiac resynchronization defibrillator (CRT-D) including telemonitoring function were enrolled. The long-term follow-up data (4.10 ± 2.58 years) were assessed. All the patients (n = 26, 100%) received daily ICD/CRT-D telemonitoring. In most of the patients (73%, n = 19), the telemedical center had to take action for a mean of three times. An acute alert due to sustained ventricular arrhythmias (VAs) occurred in 12 patients (63%) with 50% of them (n = 6) requiring ICD shock delivery. Eight patients (67%) were hospitalized due to symptomatic VAs. In 11 patients (92%), immediate medication adjustments were recommended. Relevant lead issues were revealed in thirteen patients (50%), with six patients (46%) undergoing consecutive lead revisions. Most of the events (83%) were detected within 24 h. Daily telemonitoring significantly reduced the number of in-hospital device controls by 44% (p < 0.01). The telemonitoring ensured that cardiac arrhythmias and device/lead problems were identified early, allowing pre-emptive and prompt interventions. In addition, the telemonitoring significantly reduced the number of in-hospital device controls in this cohort of HF patients.
No takes yet. Share an insight, caveat, or question.
Guckel et al. (2022) conducted a cohort in End-stage heart failure with a ventricular-assistance device (n=26). Daily implant-based multiparametric telemonitoring was evaluated on Number of in-hospital device controls (44% reduction, p=< 0.01). Daily implant-based multiparametric telemonitoring significantly reduced the number of in-hospital device controls by 44% (p < 0.01) in heart failure patients with a ventricular assist device.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:
More from Life