Key result
Arteriosonde 1010 performed better than Infrasonde 3000 for indirect blood pressure measurement, while the Pedisphyg system yielded accurate systolic values compared to intra-arterial pressure.
Why the study?
Do indirect blood pressure measuring devices (Arteriosonde 1010, Infrasonde 3000, Pedisphyg system) accurately measure systemic arterial blood pressure compared to direct intra-arterial measurement in infants and children?
Observational
Do indirect blood pressure measuring devices (Arteriosonde 1010, Infrasonde 3000, Pedisphyg system) accurately measure systemic arterial blood pressure compared to direct intra-arterial measurement in infants and children?
Among indirect blood pressure devices in infants and children, Pedisphyg accurately measures systolic pressure, while Arteriosonde and Infrasonde only approximate direct intra-arterial pressures.
Answers are grounded in this paper and cite their sources.
Pedisphyg may aid systolic BP assessment in infants and children; leaves open optimal device selection pending prospective validation.
The systemic arterial blood pressures obtained in infants and children utilizing three indirect measuring devices--Arteriosonde 1010. Infrasonde 3000, Pedisphyg system--were compared to each other and to intra-arterial pressure measured directly. The results indicate that Arteriosonde performs considerably better than Infrasonde; nevertheless, the Arteriosonde values are often only approximations of true systole and diastole. The Pedisphyg system yields accurate, reproducible values for systole; however, the system is not designed to determine diastolic blood pressure.
No takes yet. Share an insight, caveat, or question.
Reder et al. (1978) reported an observational. Indirect blood pressure measuring devices (Arteriosonde 1010, Infrasonde 3000, Pedisphyg system) vs. Direct intra-arterial pressure measurement was evaluated on Systemic arterial blood pressure accuracy. Arteriosonde 1010 performed better than Infrasonde 3000 for indirect blood pressure measurement, while the Pedisphyg system yielded accurate systolic values compared to intra-arterial pressure.