Answers are grounded in this paper and cite their sources.
Most Shack-Hartmann based aberrometers use infrared light, for the comfort of the patients. A large amount of the light that is scattered from the retinal layers is recorded by the detector as background, from which it is not trivial to estimate the centroid of the Shack-Hartmann spot. For a centroiding algorithm, background light can lead to a systematic bias of the centroid positions towards the centre of the software window. We implement a matched filter algorithm for the estimation of the centroid positions of the Shack-Hartmann spots recorded by our aberrometer. We briefly present the performance of our algorithm, and recall the well-known robustness of the matched filter algorithm to background light. Using data collected on 5 human eyes, we parameterise a simple and fast centroiding algorithm and reduce the difference between the two algorithms down to a mean residual wavefront of 0.02 microm rms.
No takes yet. Share an insight, caveat, or question.
Leroux et al. (2010) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: