Key result
The developed capacitive coupling electrode with contact feedback allows monitoring skin-electrode contact quality, revealing that contact violation leads to significant ECG signal deterioration.
Why the study?
The study aimed to develop capacitive coupling electrodes capable of monitoring skin-electrode contact quality and to experimentally identify ECG signal distortions caused by contact disturbances.
The development of a capacitive coupling electrode with contact feedback can improve the quality of ECG signal registration in personalized medicine and tele-ECG devices.
May enhance ECG quality in tele-ECG devices; hypothesis-generating and should not yet change practice.
Purpose The purpose of this paper is the development and study of capacitive coupling electrodes with the ability to monitor the quality of the skin–electrode contact in the process of electrocardiogram (ECG) diagnostics. The study’s scope embraces experimental identification of distortions contributed into the recorded ECG signal at various degrees of disturbance of the skin–electrode contact. Design/methodology/approach A capacitive coupling electrode is designed and manufactured. A large number of experiments was carried out to record ECG signals with different quality of the skin–electrode contact. Using spectral analysis, the characteristic distortions of the ECG signals in the event of contact disturbance are revealed. Findings It was found that the violation of the skin–electrode contact leads to significant deterioration in the recorded signal. In this case, the most severe distortions appear with various violations of the skin–electrode contact of two sensors in one lead. It has been experimentally shown that the developed sensor allows monitoring the quality of the contact, and therefore, improvement of the quality of signal registration, enabled by the use of bespoke processing algorithms. Practical implications These sensors will be used in personalized medicine devices and tele-ECG devices. Originality/value In this work, authors studied the effect of the skin–electrode contact of a capacitive electrode with the body on the quality of the recorded ECG signal. Based on the studies, the necessity of monitoring contact was shown to improve the quality of diagnostics provided by personalized medicine devices; the capacitive sensor with contact feedback was developed.
No takes yet. Share an insight, caveat, or question.
Евтушенко et al. (2020) studied ECG diagnostics. Capacitive coupling electrode with skin-electrode contact control was evaluated on Distortions in recorded ECG signal at various degrees of disturbance of the skin-electrode contact. The developed capacitive coupling electrode with contact feedback allows monitoring skin-electrode contact quality, revealing that contact violation leads to significant ECG signal deterioration.