Does 150-J biphasic truncated exponential shock provide equivalent first-shock defibrillation efficacy compared to 200-J monophasic damped sine shock in patients with electrically-induced ventricular arrhythmias?
Cohort (n=96)
Does 150-J biphasic truncated exponential shock provide equivalent first-shock defibrillation efficacy compared to 200-J monophasic damped sine shock in patients with electrically-induced ventricular arrhythmias?
150-J biphasic truncated exponential shocks provide highly effective defibrillation for electrically-induced VF, equivalent to 200-J monophasic shocks but with significantly lower peak current.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Higgins et al. (2004) conducted a cohort in Electrically-induced ventricular fibrillation or ventricular tachycardia (n=96). 150-joule impedance-compensated biphasic truncated exponential (BTE) shocks vs. 200-J monophasic damped sine (MDS) shocks (historical control) was evaluated on First-shock success rate for ventricular fibrillation (p=0.001 for equivalence).