Key result
Measurement of the 24-segment sphericity index of the fetal heart using FetalHQ demonstrated relatively low reliability when performed by experienced sonographers.
Why the study?
The change with advancing gestation and reproducibility of the 24-segment sphericity index of the fetal heart using FetalHQ needed evaluation.
Observational (n=81)
The 24-segment sphericity index of the fetal heart measured by FetalHQ has relatively low reliability even when performed by experienced sonographers, suggesting a need for more training on reference point placement.
Low reliability of fetal sphericity index by FetalHQ warrants caution before clinical use; leaves open need for standardized training in future validation studies.
Objective To evaluate the 24-segment sphericity index (SI) of the fetal heart using FetalHQ regarding the change with advancing gestation and reproducibility. Methods Eighty-one pregnant women at 18–21 + 6 and 28–31 + 6 weeks were studied using FetalHQ to calculate 24-segment SI of left and right ventricles. Intra- and inter-class correlation coefficients and intra- and inter-observer agreements of measurements for SI were assessed in each segment. Results With respect to intra-observer reproducibility, all SI values showed correlations between 0.596 and 0.774 for the left ventricle, and between 0.491 and 0.764 for the right ventricle, with good intra-observer agreements except for right ventricular segment 11–15. With respect to inter-observer reproducibility, all SI values showed correlations between 0.473 and 0.755 for the left ventricle, and between 0.48 and 0.767 for the right ventricular segment 1–18. However, right ventricular segment 19–24 showed no correlations. Good inter-observer agreements were noted in the right ventricle, whereas poor inter-observer agreements were evident in the left ventricle. There were significant differences in the mean SI values in the apical segment of the left ventricle between 18–21 + 6 and 28–31 + 6 weeks (p < .05), whereas there were no significant differences in the mean SI values in the right ventricle between both gestational ages. Conclusion These results show that measurements of 24-segment SI of fetal heart had relatively low reliability of measurements performed by experienced sonographers. This may suggest that more training on proper placement of the 3 end-systolic and end-diastolic reference points during the initial analysis may be relevant.
No takes yet. Share an insight, caveat, or question.
Hata et al. (2020) conducted an observational in Pregnancy (fetal heart evaluation) (n=81). FetalHQ 24-segment sphericity index measurement was evaluated on Intra- and inter-observer reproducibility and agreements of 24-segment sphericity index measurements. Measurement of the 24-segment sphericity index of the fetal heart using FetalHQ demonstrated relatively low reliability when performed by experienced sonographers.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: