Recently a number of glucose dehydrogenase-based point of care (POCT) systems for glucose monitoring were successfully introduced on the market. Icodextrin, a glucose polymer is widely used as an osmotic agent in continuous ambulatory peritoneal dialysis (CAPD). Its metabolites are substrates for glucose dehydrogenase, inducing an analytical error which is gaining importance in the determination of glucose in a hospital environment. Since icodextrin is hydrolysed by amylase in the extracellular fluids, the analytical error in vivo is aggravated by the presence of the smaller oligosaccharides which originate from amylase activity. Clinicians should be warned about the spurious high glucose results which might occur in icodextrin-treated patients. In particular in in conditions associated with increased amylase activities, analytical errors are to be expected in icodextrin-treated patients. Alternative glucose determination methods should be recommended in the latter group of patients.
No takes yet. Share an insight, caveat, or question.
Floré et al. (2006) studied this question.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: