Answers are grounded in this paper and cite their sources.
A high-precision temperature control method using flowing water near room temperature is described in this paper. The temperature of a thermostatic cell immersed in a water flow within a water jacket is regulated by controlling the temperature of the water circulating through the jacket. This control is accomplished by three-step controlled heating and constant cooling. The method provides control with a short response time. Temperature stabilization-process control within a fluctuation of ±25 μK over a period of 12 h and temperature variation-process control at a rate of 320 μK/h with a fluctuation of less than ±50 μK are achieved. All of the electric devices for controlling and measuring, as well as the thermostatic cell, are placed in thermostatic containers the temperature of which is controlled within a fluctuation of ±0.05 K. The total error in these temperature measurements is estimated to be approximately ±11 μK.
No takes yet. Share an insight, caveat, or question.
H. Ogasawara (1986) studied this question.