Key result
The real-time heart rate measurements using the electronic stethoscope system differed from manual measurements by only 2-5%, with a delay of 1.7 to 2 seconds.
Why the study?
More accurate stethoscope measurements are needed to diagnose heart problems earlier, monitor patients, and provide initial clinical data for physicians.
Does a low-cost electronic stethoscope with a LabVIEW interface accurately detect real-time heart rate compared to manual measurement?
Does a low-cost electronic stethoscope with a LabVIEW interface accurately detect real-time heart rate compared to manual measurement?
A low-cost electronic stethoscope prototype utilizing a 4th-order Butterworth filter and LabVIEW interface can detect real-time heart rate with a 2-5% error margin compared to manual measurement.
Answers are grounded in this paper and cite their sources.
May facilitate low-cost real-time monitoring; leaves open clinical validation in larger cohorts.
Better accurate stethoscope measurements have been needed to diagnose heart problems earlier, monitor patients, and provide initial clinical data for physicians. This study aims to evaluate an electronic stethoscope for automatic identification of heart rate by monitoring Beats Per Minute (BPM) in real-time. In this work, a new design with a low cost electronic stethoscope is designed and implemented as a simple circuit for a replacement with conventional stethoscopes. This presented a low cost stethoscopes consisting of a preamplifier, low-pass filter, high-pass filter, microcontroller and Bluetooth. This simulation and experimental study was carried out for electronic circuit design testing. The condenser microphone transmited the signal into the signal conditioning circuit, and amplified it from 10 to 30 times. The low- and high- pass filter circuit was with cutoffs of 180 Hz and 50 Hz, respectively. The result shows that the fourth-order Butterworth filter was the best filter, with a gain of 0.707 Volt, -3.01 dB, and 0.782 Volt, -2.137 dB, respectively. The real-time measurements using the system are not significantly different from the manual measurements, with around 2-5%, with a delay of 1.7 to 2 seconds. The results indicated that an electronic stethoscope system could provide suitable information for BPM value and could display heart sound signals.
No takes yet. Share an insight, caveat, or question.
Oktivasari et al. (2020) studied Healthy volunteers (n=8). Electronic stethoscope with LabVIEW vs. Manual measurement was evaluated on Difference in Beats Per Minute (BPM) compared to manual measurement (2-5% difference). The real-time heart rate measurements using the electronic stethoscope system differed from manual measurements by only 2-5%, with a delay of 1.7 to 2 seconds.