Key result
Harmonic imaging during transthoracic echocardiography improved sensitivity (100% vs 78%) but decreased specificity (82% vs 100%) for detecting patent foramen ovale compared to fundamental imaging.
Why the study?
Does harmonic imaging during transthoracic echocardiography improve the diagnostic accuracy for detecting patent foramen ovale compared to fundamental imaging?
Cross-Sectional (n=64)
Does harmonic imaging during transthoracic echocardiography improve the diagnostic accuracy for detecting patent foramen ovale compared to fundamental imaging?
Absolute Event Rate: 100% vs 78%
Harmonic imaging during transthoracic echocardiography improves sensitivity but reduces specificity for PFO detection compared to fundamental imaging, largely due to transpulmonary transit of bubbles.
Harmonic imaging trades PFO sensitivity gains for specificity losses; leaves open optimal TTE protocol pending validation studies.
UNLABELLED: Contrast echocardiography using agitated saline is widely employed to detect right to left shunt in patients with suspected patent foramen ovale (PFO). The sensitivity of fundamental transthoracic echocardiography (FTTE) is often limited by poor far-field resolution. The purpose of this study was to evaluate the test characteristics of harmonic imaging during transthoracic echocardiography (HTTE) for the detection of PFO. METHODS: Sixty-four patients referred for transesophageal echo (TEE) for clinical indications underwent FTTE and HTTE during saline contrast injections. RESULTS: TEE identified nine PFOs. For FTTE, the sensitivity was 78% and the specificity was 100%. For HTTE, the sensitivity was 100% and the specificity was 82%. All ten false positives with HTTE occurred after five cardiac cycles, suggesting transpulmonary transit of the bubbles. CONCLUSION: In the detection of PFO, HTTE improves sensitivity at the expense of specificity, particularly if the timing of contrast appearance in the left heart is ignored.
No takes yet. Share an insight, caveat, or question.
Madala et al. (2004) conducted a cross-sectional in Suspected patent foramen ovale (PFO) (n=64). Harmonic imaging during transthoracic echocardiography (HTTE) vs. Fundamental transthoracic echocardiography (FTTE) was evaluated on Sensitivity for the detection of PFO. Harmonic imaging during transthoracic echocardiography improved sensitivity (100% vs 78%) but decreased specificity (82% vs 100%) for detecting patent foramen ovale compared to fundamental imaging.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: