Answers are grounded in this paper and cite their sources.
The fretting corrosion resistances of two alloys that are currently used in surgical implants (Type 316L stainless steel and cast Vitaliium alloy) have been compared with the resistances of two high strength, highly ductile, metastable austenitic stainless steels (TRIP steels). The fretting corrosion apparatus used allowed control of the environmental composition, the vibrational amplitude, the vibrational frequency, and the stress between the two metal surfaces in fretting contact. The rate of corrosion of the contacting surfaces during fretting was significantly higher than the rate before or after the fretting period, and the corrosion rate during fretting generally increased with time. Corrosion rate data taken after 30 minutes of fretting indicated a linear dependence on amplitude, frequency, and contact stress. Cast Vitallium alloy appeared to have a somewhat higher resistance to fretting corrosion than the other three alloys. The available data suggests that fretting increases the corrosion rate because the passive surface film is repeatedly disrupted and repaired. However, the rate of corrosion appears to be controlled by the rate of oxygen transport from the bulk solution to the cathodic sites on the metal surface.
No takes yet. Share an insight, caveat, or question.
Syrett et al. (1978) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: