Key result
The Glucometer 4 blood glucose system demonstrated within-batch precision of 1.1-4.2% and between-batch precision of 2.6-6.7% coefficient of variation, making it suitable for near-patient testing.
Why the study?
Does the Glucometer 4 blood glucose system provide precise near-patient testing compared to the reference method in an out-patient diabetes clinic?
Observational
No
Does the Glucometer 4 blood glucose system provide precise near-patient testing compared to the reference method in an out-patient diabetes clinic?
The Glucometer 4 blood glucose system demonstrated acceptable precision and was deemed suitable for near-patient testing by nurses and patients.
The performance of a new system for near‐patient monitoring of blood glucose levels based on the reference hexokinase/glucose‐6‐phosphate dehydrogenase method was evaluated in a routine out‐patient diabetes clinic. The system includes features designed to overcome operator dependency of results. Within‐batch precision was 1.1‐4.2% coefficient of variation, while between‐batch coefficients of variation of 2.6‐6.7% were achieved. The new system was assessed to be suitable for use by nurses and patients.
No takes yet. Share an insight, caveat, or question.
MP Powell (1994) conducted an observational in Diabetes. Glucometer 4 blood glucose system was evaluated on Precision (coefficient of variation). The Glucometer 4 blood glucose system demonstrated within-batch precision of 1.1-4.2% and between-batch precision of 2.6-6.7% coefficient of variation, making it suitable for near-patient testing.