The International Atomic Energy Agency has long been aware of the need for improved dosimetric accuracy in the medical applications of radiation, such as in teleradiotherapy.Of all radiotherapy centres in the world, probably less than 50 % have ready access to standardizing dosimetry laboratories; and even these laboratories at present provide only standards of radiation exposure (in roentgen), whereas the decisive quantity in radiotherapy is the absorbed dose (in rad) received by the tissues being irradiated.I n order to assist radiotherapy centres, particularly in the developing countries, with the determination of absorbed dose in 6"Co teletherapy, the IAEA has set up a postal dose-intercomparison service utilizing thermoluininescent LiF powder.Prior to putting this scheme into practice, the intercomparison procedures were tested with the co-operation of 19 radiotherapy centres in 6 countries.The participating institutes are listed in Table 1.This report gives a brief description of the procedures and of thc results, without identifying the results with any particular institute.Method.Each participant was provided with a plastic water tank in which the irradiation was to take place, as well as with instructions for the experimental