An open-ended co-axial probe was designed for monitoring density changes in low-permittivity fluids. The design was based on the Marcuvitz equation for the probe admittance and was tested with measurements on liquids of known permittivities at room temperature and atmospheric pressure. A simple calibration procedure and a bilinear calibration procedure were tested. Only the bilinear procedure was sufficiently accurate for monitoring density changes. The reproducibility of the measured high-frequency permittivity was found to be 0.005 at room temperature and atmospheric pressure. From measurements on mixtures of reference chemicals it was found that density differences of could be measured. The system was also used to measure the permittivity and density changes in the temperature range 20 - C and the pressure range 1 - 150 bar for n-decane, n-heptane and a petroleum-gas condensate. At temperatures above room temperature the reflection coefficients of liquids under test were corrected with a linear model. This accounted for changes in the measurement system. The results agreed with the temperature-induced density changes given in the literature. The small pressure-induced density changes could not be detected. This indicated that the probe can be used to monitor temperature-induced density changes in low-permittivity liquids when a bilinear calibration procedure and a linear temperature correction are used.