Key result
EWA, FWHM, and FACT algorithms performed on par with manual delineation for quantifying infarct size, with biases of -0.48±3.1%, -0.3±4.4%, and 2.3±4.2% of the left ventricle, respectively.
Why the study?
Many algorithms quantify infarct size in LGE images, but few have been benchmarked against ex-vivo measurements, and reported performance varies considerably across studies.
Several automated algorithms (EWA, Heiberg-08, FWHM, FACT) perform comparably to manual delineation for quantifying infarct size on LGE CMR, highlighting the need for consistent methodology in clinical trials.
LGE infarct algorithms vary in accuracy; leaves open the optimal method for clinical use pending standardized validation.
BACKGROUND: In cardiovascular magnetic resonance, late gadolinium enhancement (LGE) is the standard method to visualize myocardial infarction (MI). Many algorithms quantifying infarct size in LGE images exist. However, only few algorithms have been validated, i.e., benchmarked against an ex-vivo measurement. Furthermore, the reported algorithm performance varies considerably between studies. OBJECTIVES: The aim of this study was to compare the performance of all infarct measurement algorithms against an ex-vivo measurement and to promote a discourse regarding advantages and disadvantages of individual measurement methods. METHODS: MI was induced in 22 pigs. In-vivo LGE imaging was conducted on d0, d3 or d7 post-MI. For ex-vivo validation infarct was measured using high-resolution T1-weighted images. In-vivo infarct size was measured using the full-width at half-maximum (FWHM), n-SD from remote (2,3,5, and 6 SD), feature analysis and combined thresholding (FACT), expectation maximization-weighted A priori information (EWA), Heiberg-08 and Otsu algorithms and manual delineation. No manual adjustments were made to algorithm delineations. RESULTS: Clear differences in variance and bias were observed between algorithm-based methods, and no method performed optimally in this heterogeneous dataset where the best had a bias of -0.48±3.1, -0.3±4.4%, 2.3±4.2% left ventricle for EWA, FWHM, and FACT, respectively. Manual delineation by experienced observers performed well with a bias of 1.9±5.4%. CONCLUSION: EWA, Heiberg-08, FWHM, and FACT all perform on par with manual delineation, however, Heiberg-08, and FWHM are not suitable for phase sensitive inversion recovery images. The technique used to measure infarct size should be disclosed in clinical trials and in original research. Caution should be applied when comparing datasets employing different infarct quantification methods. Manual infarct delineation by experienced readers remains a reliable technique to measure infarct size.
No takes yet. Share an insight, caveat, or question.
Kopic et al. (2025) studied Myocardial infarction (n=22). Infarct measurement algorithms (EWA, FWHM, FACT, etc.) vs. Ex-vivo measurement and manual delineation was evaluated on Bias in infarct size measurement compared to ex-vivo measurement. EWA, FWHM, and FACT algorithms performed on par with manual delineation for quantifying infarct size, with biases of -0.48±3.1%, -0.3±4.4%, and 2.3±4.2% of the left ventricle, respectively.