The precise determination of the intensity I of a source of radiation is often accomplished by taking measurements at several distances from the source and relating them through the inverse square law, I = Vr2, where I is related to the intensity of the source through the constants of the experiment and V is the response of a detector at a distance r from the source (Mayneord and Roberts, 1937; Ghosh, Kastner, and Whyte, 1953). In general, the detector output measurements are more accurate close to the source whereas the measurements of r are more accurate at large distances. The latter is especially true if the source or detector has dimensions that are not small compared to r. A method has been devised that is independent of the accurate location of the effective centres of source and detector, thus permitting precise measurements to be made at much smaller distances than were previously possible. The method also analytically determines the difference between the assumed and actual separation between the effective centres of the source and detector. Consider a point detector constrained so that it can only move along an arbitrary straight line. Let b be the perpendicular distance from a point source of radiation to this line, and a the distance along the line from the intersection of b to an arbitrary zero point O from which the distance d to the detector is measured (Fig. 1).