Answers are grounded in this paper and cite their sources.
The paper deals with problems of percentage depth dose, gamma-ray filters, and the measurement of gamma rays in r units. The results of measurements in a wax phantom for a gamma-ray source at F.S.D. from 5 to 75 cm. are discussed. It is pointed out how low are the values of D10 obtainable with gamma rays used at short distances from the skin compared to those from high voltage X rays. At comparable F.S.D., however, the gamma-ray source gives values considerably in excess of those obtained at 400 KV. The second section of the paper shows the advantages of the use of heavy element filters in surface therapy. The last section brings forward considerable additional evidence for the value of the ionisation current observed by the writer in a small air-wall chamber under the action of a known quantity of Ra (B+C). It is pointed out that this value is in accord with recent determinations of Eve's number by a series of workers on cosmic rays, as well as investigations of L. H. Gray, of the total ionisation produced on a scattering medium by radium. All these results diverge greatly from results of parallel plate chamber measurements by Failla and Henshaw, who obtained values of the ionisation 4·5 times smaller that the air-wall chamber results. It is pointed out, however, that the parallel plate chamber results of Bruzau are not in accord with Failla and Henshaw, but are more nearly in agreement with the results of air-wall chamber experiments.
No takes yet. Share an insight, caveat, or question.
W. V. Mayneord (1933) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: