Key result
Transcatheter closure using the Cardi-O-Fix plug achieved successful device implantation and complete echocardiographic closure at 1 year in all 5 patients with muscular ventricular septal defects, with no major complications.
Why the study?
The study was conducted to evaluate the safety and efficacy of the Cardi-O-Fix plug for the treatment of muscular ventricular septal defects.
Does transcatheter closure using the Cardi-O-Fix plug safely and effectively treat muscular ventricular septal defects in pediatric patients?
Observational (n=5)
No
Does transcatheter closure using the Cardi-O-Fix plug safely and effectively treat muscular ventricular septal defects in pediatric patients?
Transcatheter closure of muscular ventricular septal defects using the Cardi-O-Fix plug appears feasible, safe, and effective at 1-year follow-up in a small pediatric cohort.
Answers are grounded in this paper and cite their sources.
Limited retrospective data indicate procedural feasibility; leaves open the role of Cardi-O-Fix plug pending prospective trials.
OBJECTIVE: The aim of this study was to evaluate the safety and efficacy of the Cardi-O-Fix plug used for the treatment of muscular ventricular septal defects. METHODS: We retrospectively reviewed the medical records of five patients with muscular ventricular septal defects who underwent transcatheter closure using the Cardi-O-Fix Plug, from November 2017 to August 2019. The median age was 5.1 years (range: 3.2-6.5). Their median body weight was 18.1 kg (range: 13.4-21.8). All the patients underwent detailed two-dimensional Doppler and colour flow imaging by transthoracic echocardiography. The left ventricular median defect size of the muscular ventricular septal defects was 5.6 mm (range: 5.3-7.0). The right ventricular median defect size of the muscular ventricular septal defects was 3.9 mm (range: 3.3-4.7). All the procedures were performed on beating hearts. RESULTS: All the patients underwent successful device implantation with no displacement or detachment, they have complete echocardiographic closure at the 1-year follow-up. There were no occluder-related arrhythmia, chordae tendineae rupture, tricuspid insufficiency, aortic regurgitation, haemolysis, or embolisation. CONCLUSIONS: Application of the Cardi-O-Fix plug appears to be a feasible, safe, and effective treatment option for patients with muscular ventricular septal defects. Longer follow-up periods are warranted to prove the conclusion for long-term outcomes.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2021) conducted an observational in Muscular ventricular septal defects (n=5). Cardi-O-Fix plug was evaluated on Successful device implantation and complete echocardiographic closure. Transcatheter closure using the Cardi-O-Fix plug achieved successful device implantation and complete echocardiographic closure at 1 year in all 5 patients with muscular ventricular septal defects, with no major complications.
Synapse has enriched 10 related papers. Consider them for comparative context: