This paper deals with the theoretical and practical investigation of a test method using superconducting cavity and helical resonators in an oscillator loop, which aIIows precision measurements to be performed on solid dielectrics in the range of 0.1-10 GHz, and below 9 K. The underlying formulas are an extension of the well-known perturbation formalism and are not restricted to low temperatures. Our experiments resulted in unloaded quality factors of Q = 5/spl dot/ 10/sup 7/, between 0.2 and 10 GHz, with a maximum Q (2.2 K, 0.19 GHz) of 9/spl dot/ 10/sup 8/, which enabled us to observe the smallest loss tangent so far: tan delta (2.2 K, 6.5 GHz) = (3.7 /spl plusmn/ 5 percent)/spl dot/ 10/sup -7/ in polyethylene.
No takes yet. Share an insight, caveat, or question.
W. Meyer (1977) studied this question.