Key result
Real-time 3D echocardiography significantly underestimated small volumes by 6.7% compared to 13.1% with 2D echocardiography, with image acquisition variables heavily impacting accuracy (P<0.001).
Why the study?
Does real-time 3D echocardiography improve the accuracy of small volume, area, and distance measurements compared to 2D echocardiography in phantoms and neonates/infants?
Does real-time 3D echocardiography improve the accuracy of small volume, area, and distance measurements compared to 2D echocardiography in phantoms and neonates/infants?
Absolute Event Rate: -6.7% vs -13.1%
p-value: p=<.05
Real-time 3D echocardiography is a reliable method for calculating small distances, areas, and volumes comparable to neonatal hearts, but requires standardized protocols to minimize acquisition and analysis errors.
OBJECTIVES: The aim of this study was to assess the validity, accuracy, and reproducibility of real-time 3-dimensional (3D) echocardiography for small distances, areas, and volumes. METHODS: Real-time 3D echocardiography using matrix technology was performed in small calibrated tissue-mimicking phantoms and compared with 2-dimensional (2D) echocardiography. In a systematic variation of variables on data acquisition and analysis including different 3D workstations (manual disk summation versus semiautomatic border detection), the relative contributions of sources of errors were determined. The clinical relevance of the in vitro findings was assessed in 5 neonates and infants. RESULTS: Distance calculation was valid (mean relative error ± SD, -0.15% ± 1.2%). Underestimation of areas and volumes was significant for both 2D and 3D echocardiography (area: 2D, -7.0% ± 2.9%; 3D, -6.0% ± 2.8%; volume: 2D, -13.1% ± 4.5%; 3D, -6.7% ± 2.5%; P < .05). Adjustment of compression and gain on data acquisition (difference of the means: 2D, 11.6%; 3D, 17.9%), gain on postprocessing (3D, 3.4%), and the border detection algorithm on analysis (2D, 4.8%; 3D, 16.6%) had a highly significant effect on volume and area calculations (P < .001). In vivo, compression and gain on acquisition (3D, 19.1%) and the 3D workstation on analysis (3D, 22.2%) had a highly significant impact on left ventricular volumetry (P < .001). CONCLUSIONS: Real-time 3D echocardiography is a reliable method for calculation of small distances, areas, and volumes comparable with the size of the neonatal and infant heart. Variables influencing boundary identification during image acquisition and analysis have a significant impact on 2D and 3D area and volume calculations. Standardized protocols are mandatory to avoid these sources of error in both clinical practice and research.
No takes yet. Share an insight, caveat, or question.
Herberg et al. (2011) studied Neonates and infants requiring echocardiography (n=5). Real-time 3-dimensional (3D) echocardiography vs. 2-dimensional (2D) echocardiography was evaluated on Underestimation of volume (p=<.05). Real-time 3D echocardiography significantly underestimated small volumes by 6.7% compared to 13.1% with 2D echocardiography, with image acquisition variables heavily impacting accuracy (P<0.001).