Answers are grounded in this paper and cite their sources.
The disinfection procedures aim to reduce the microbial load, but there are doubts about the risks of contamination spreading into the lumens of devices, such as three-way stopcocks (3-WS). This study aimed at an in vitro evaluation of the antibacterial procedure of manual friction of 3-WS intentionally contaminated and to determine the solution dispersion into the lumens. Laboratory experiments were developed in two steps: evaluation of bacterial spread through intentional contamination with Staphylococcus aureus and Pseudomonas aeruginosa, and alcohol dispersion into the 3-WS lumens. After manual friction of the 3-WS with saline solution at 0.85% (w/v) [control group], S. aureus and P. aeruginosa were isolated in the lumens of 55.6% and 27.8% of the devices, respectively. However, after the disinfection of the 3-WS with ethyl alcohol at 70% (w/v), there was no bacterial contamination in the lumens of the 3-WS. On the other hand, the solution dispersion (dye) into the lumens was evidenced by two different techniques: Durham tubes (5.6%) and swabs (46.3%). The manual friction of the 3-WS with ethyl alcohol at 70% demonstrated antibacterial efficacy, but it refers to reflections on the risk of solution diffusion into the venous network and the inherent clinical practice situations and patient safety.
No takes yet. Share an insight, caveat, or question.
Souza et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: