Key result
Left atrio-femoral bypass with a Bio-pump and evoked spinal potential monitoring during elective resection of descending thoracic or thoraco-abdominal aortic aneurysms resulted in no cases of renal failure or spinal cord injury among 6 patients.
Why the study?
Does left atrio-femoral bypass with Bio-pump and ESP monitoring prevent renal failure and paraplegia in patients undergoing elective resection of descending thoracic or thoraco-abdominal aortic aneurysm?
Observational (n=6)
No
Does left atrio-femoral bypass with Bio-pump and ESP monitoring prevent renal failure and paraplegia in patients undergoing elective resection of descending thoracic or thoraco-abdominal aortic aneurysm?
Left atrio-femoral bypass with a Bio-pump and ESP monitoring is a safe and effective adjunct to prevent renal and spinal cord ischemic complications during elective descending thoracic or thoraco-abdominal aortic aneurysm resection.
Answers are grounded in this paper and cite their sources.
Supports feasibility in small elective series; leaves open whether it reduces complications in larger prospective studies.
During a 8-month period (Apr. -Nov. 1988), we underwent 6 elective resections of descending thoracic or thoraco-abdominal aneurysm. In all cases, left atrio-femoral bypass with Bio-pump was used during the cross clamping of the thoracic aorta. Prior to the cross clamping, heparin (0.5-1.0mg/kg) was administrated intravenously to keep the activated clotting time more than 200sec, then the pump was turned on. The bypass with the pump was used for 90 to 240min, and the mean flow rate ranged 1000 to 2400ml/min to maintain mean arterial pressure of the dorsal pedis artery above 60mmHg. These flow rates were enough to reserve urin volume during the bypass, and leading to no renal failure. To keep the body temperature from cooling down we made a useful bypass circuit with a simple heat exchanger, in which the usual blanket warmer was the heat source. Evoked spinal potential was monitored to prevent the patient from paraplesia. There were some changes in the wave in 3 cases out of 6. Two of the 3 showed temporary changes during declamping shock, and the remaining resulted in a permanent drop of voltage. But none of them had any evidence of spinal cord injury.
No takes yet. Share an insight, caveat, or question.
Nishikimi et al. (1990) conducted an observational in Descending thoracic or thoraco-abdominal aortic aneurysm (n=6). Left atrio-femoral bypass with Bio-pump and evoked spinal potential monitoring was evaluated on Spinal cord injury (paraplegia) and renal failure. Left atrio-femoral bypass with a Bio-pump and evoked spinal potential monitoring during elective resection of descending thoracic or thoraco-abdominal aortic aneurysms resulted in no cases of renal failure or spinal cord injury among 6 patients.