Answers are grounded in this paper and cite their sources.
To develop a simple micro-platform for gene mutation detection, we used a DNA hybridization approach in a combined microchannel. A glass groove immobilized with oligonucleotide probes by UV-crosslinking, and encased in a transparent heat-shrinkable polythene tube. After heat treatment, the polythene tube stretched tightly over the glass groove and was designated as a combined capillary chip (CCC). The CCC assay was optimized with 10 μL reaction solutions shuttling back and forth at 50 μL min−1 for 10 min, give a detection minimum of 0.1 nM target DNA sequences. In hepatitis B virus (HBV) YMDD mutations detection, 61.3% (92/150) was of a single genotype and it was authenticated with sequence analysis. The other 38.7% (58/150) was mixed genotype detected, but 18.9% (11/58) has a missed diagnosis in sequence analysis. It proved a higher sensitivity than sequence analysis. The CCC assay has a simple fabricating process, simple structure and higher specificity in gene mutation detection. These features are promising for clinical gene mutation analysis.
No takes yet. Share an insight, caveat, or question.
An et al. (2012) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: