Key result
Valve-in-valve transcatheter aortic valve replacement successfully reduced the bioprosthetic AV peak systolic pressure gradient from 90 to 29 mmHg and average gradient from 42 to 19 mmHg.
Case Report (n=1)
Valve-in-valve transcatheter aortic valve replacement is a safe and effective minimally invasive option for treating structural degeneration of a previously implanted bioprosthetic aortic valve in high-risk surgical patients.
Supports valve-in-valve TAVI feasibility in high-risk bioprosthetic degeneration; leaves open durability, selection, and generalizability questions.
Introduction. We present the clinical observation of a 72-year-old female patient with high surgical risk and structural degeneration of a bioprosthetic aortic valve (AV) cusps in the form of stenosis, accompanied by severe dysfunction. Transcatheter implantation of bioprosthesis Medtronic CoreValve™ Evolut™ R-23 was performed using the valve-in-valve technique. The choice of minimally invasive treatment tactics is substantiated, a preoperative examination algorithm and a specific bioprosthesis model for such intervention are provided. Materials and methods. Imaging – echocardiography (Echo), electrocardiography, multispiral computed tomography, coronary angiography. Bioprosthetic valve calcification and stenosis with critical parameters of the bioprosthetic AV peak pressure gradient according to Echo data were the indications for minimally invasive surgery. Results. Dynamic observation revealed a progressive deterioration in the function of the previously implanted bioprosthetic heart valve in the aortic position, and a critical deterioration in the patient’s condition. After additional examination of the patient and selection of a new prosthesis, valve-in-valve transcatheter aortic valve replacement was done. The positive dynamics of the general state of the patient was noted in the early postoperative period. Echo data showed that the bioprosthetic AV peak systolic pressure gradient decreased from 90 to 29 mmHg, average gradient – from 42 to 19 mmHg. Conclusion. The minimally invasive valve-in-valve transcatheter aortic valve replacement used to correct the dysfunction of a bioprosthetic AV that was previously implanted during an open surgery was shown to be safe and effective and can be considered as one of the options for repeat valve replacement.
No takes yet. Share an insight, caveat, or question.
Коков et al. (2020) conducted a case report in Structural degeneration of a bioprosthetic aortic valve (n=1). Valve-in-valve transcatheter aortic valve replacement (Medtronic CoreValve Evolut R-23) was evaluated on Bioprosthetic AV peak systolic pressure gradient and average gradient. Valve-in-valve transcatheter aortic valve replacement successfully reduced the bioprosthetic AV peak systolic pressure gradient from 90 to 29 mmHg and average gradient from 42 to 19 mmHg.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: